Short-term outcomes of transcatheter atrial septal defect closure without balloon sizing at Hanoi Heart Hospital

Quoc Hung Nguyen1, , Thi Hai Van Dang, Toan Thang Nguyen, Dac Long Tran, Cong Ha Nguyen
1 Hanoi Heart Hospital

Main Article Content

Abstract

Background: Balloon sizing is a conventional method used to determine an appropriate device size for percutaneous closure of atrial septal defects (ASDs). However, balloon sizing has certain limitations, including potential overestimation of the actual defect size, increased treatment costs, and prolonged procedure time. Objective: To evaluate the short-term outcomes of percutaneous ASD closure without balloon sizing at Hanoi Heart Hospital. Methods: From October 2025 to June 2026, 104 patients with ASD underwent percutaneous device closure without balloon sizing at Hanoi Heart Hospital. Clinical, echocardiographic, procedural, and post-procedural follow-up data were collected and analyzed. The mean follow-up duration was 5.6 months, ranging from 3 to 10 months. Results: Procedural success was achieved in all 104 patients (100%), defined as successful deployment of the device in the appropriate position, stable device fixation on testing, and absence of compression or obstruction of adjacent anatomical structures. In 98 patients (94.2%), the device was successfully deployed using the initially selected size according to the sizing formula. The remaining 6 patients (5.8%) required adjustment of device size or balloon-assisted deployment: 2 required a larger device, 2 required a smaller device, and 2 required balloon assistance. All 6 cases were successfully completed. No vascular complications, device embolization, or pericardial effusion were observed during follow-up. Three patients (2.9%) developed arrhythmias: 2 had sinus bradycardia and 1 developed transient third-degree atrioventricular block one week after the procedure. All arrhythmias resolved completely without the need for permanent pacemaker implantation. Conclusion: Percutaneous ASD closure without balloon sizing is a safe and effective treatment option for patients with suitable anatomical characteristics, with a high procedural success rate and short procedure and fluoroscopy times. This approach may reduce the use of additional consumable materials during the procedure. However, further studies with larger sample sizes and longer follow-up are needed to evaluate mid- and long-term outcomes and to identify factors associated with technical failure.

Article Details

References

1. Hoffman JIE, Kaplan S. The Incidence of Congenital Heart Disease. J Am Coll Cardiol. 2002; 39:1890–1900.
2. King TD, Thompson SL, Steiner C. Secundum atrial septal defect. Nonoperative closure during cardiac catheterization. JAMA. 1976; 235:2506–2509.
3. Du ZD, Hijazi ZM, Kleinman CS. Amplatzer Investigators: Comparison between transcatheter and surgical closure of secundum atrial septal defect in children and adults: results of a multicenter nonrandomized trial. J Am Coll Cardiol. 2002; 39:1836–1844.
4. Rosas M, Zabal C, Garcia-Montes J. Transcatheter versus surgical closure of secundum atrial septal defect in adults: impact of age at intervention. A concurrent matched comparative study. Cong heart dis. 2007; 2:148–155.
5. Harper RW, Mottram PM, McGaw DJ. Closure of secundum atrial septal defects with the Amplatzer septal occluder device: Techniques and problems. Catheter Cardiovasc Interv 2002;57: 508–524.
6. Du ZD, Cao Q, Rhodes J. Choice of size and results of transcatheter closure of atrial septal defect using the Amplatzer septal occluder. J Interv Cardiol 2002;15: 287–292.
7. Writing Committee Members , Michelle Gurvitz, Eric V, Kriger, et al. 2025 ACC/AHA/HRS/ISACHD/SCAI Guideline for the Management of Adults With Congenital Heart Disease: A Report of the American College of Cardiology/American Heart Association Joint Committee on Clinical Practice Guidelines. Circulation Volume 153, Issue 8, 24 February 2026; Pages e115-e251.
8. Alain Fraisse1, Monica Latchman1, Shiv-Raj Sharma. Atrial septal defect closure: indications and contra-indications. J Thorac Dis 2018;10 (Suppl 24): S2874-S2881.
9. Jou-Kou Wang, Shen-Kou Tsai, Shu-Man Lin. Transcatheter Closure of Atrial Septal Defect Without Balloon Sizing. Catheterization and Cardiovascular Interventions 2008; 71:214–221.
10. Amin Z, Hijazi ZM, Bass JL, et al. Erosion of secundum atrial septal defects: Review of registry of complications and recommendations to minimize future risk. Catheter Cardiovasc Interv 2004;63:496–502.
11. Carlson KM, Justino H, OBrien RE, et al. Transcatheter atrial septal defect closure: Modified balloon sizing technique to avoid overstretching the defect and oversizing the Amplatzer septal occluder. Catheter Cardiovasc Interv 2005;66:390–396.
12. Fisher G, Kramer H, Stieh J. Transcatheter closure of secundum atrial septal defects with the new self-centering Amplatzer septal occluder. Eur Heart J 1999; 83: 933–936.
13. Walsh KP, Maadi IM. The Amplatzer septal occluder. Cardiol Young 2000; 10:493–501.
14. Rao PS, Langhough R. Relationship of echocardiographic, shunt flow, and angiographic size to the stretched diameter of the atrial septal defect. Am Heart J 1991;122:505–508
15. Godart F, Rey C, Francart C. Twodimensional echocardiographic and color Doppler measurement of atrial septal defect and comparison with the balloon-stretched diameter. Am J Cardiol 1993;72:1095–1097.
16. Ali Eshraghi, Ehsan Aghaei-Moghadam, Aliakbar Zeinaloo. Atrial Septal Defect Device Closure. Is Balloon Sizing Necessary? Iran J Pediatr. 2021 April; 31(2):e108076.
17. Beom Joon Kim, Jinyoung Song, June Huh. Outcomes After Transcatheter Closure of Atrial Septal Defect Without Using a Balloon-Sizing Technique: A Randomized Controlled Comparison with Closures Using a Balloon-Sizing Technique. The Heart Surgery Forum 2022-4571. 25 (2).
18. Saurabh Kumar Gupta, S Sivasankaran, S Bijulal. Trans-catheter closure of atrial septal defect: Balloon sizing or no balloon sizing – single centre experience. Annals of Pediatric Cardiology 2011 Vol 4 Issue 1: 28-33.
19. Hiroko Asakai, Sofia Weskamp, Lucas Eastaugh, et al. Atrioventricular block after ASD closure. Heart Asia 2016; 8: 26–31.
20. Kenji Suda, Marie-Jose´e Raboisson, Eric Piette, et al. Reversible Atrioventricular Block Associated With Closure of Atrial Septal Defects Using the Amplatzer Device. JACC Vol. 43, No. 9, 2004: 1677-82.
21. Alsoufi Mahmoud. Heart block after ASD device closure. J Cardiol Curr Res. 2014;1(7):196-198.